S1200 ICU Ventilator
Features
- l 15” TFT LCD touch screen displays the ventilation parameters, alarm information and waveforms.
- Air compressor runs stably with low noise, creates quiet environment for doctors and patients.
- Integration breathing circuit design.
- Multiple working modes such as volume control and pressure limit.
- Multiple parameters monitoring interface to show clear parameter to the operator, allow operator know the patient conditions in all aspects.
- Real time pressure-time, flow-time loop graphics, pressure volume loop graphics and high precision ETCO2, O2 concentration detection function included.
- Electrical PEEP control
- Breathing valve is disassembly and can be disinfected.
Safety
- Three levels alarm system, visual and sound alarm information.
- With multiple type of alarms, reminding and protection functions.
- Self-check before running, eliminate system malfunction.
- Advanced power management control technology.
- With built-in backup power supply, when normal power supply shut down, back-up power supply will start work automatically.
- Separate design of electric and gas, keep safety running of ventilator.
Specifications
| Ventilation modes: | IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV |
| Ventilator parameter ranges | |
| Tidal volume(Vt) | 0 ~ 2000mL |
| Frequency (Freq)
Oxygen concentration |
1bpm ~ 100 bpm
21%-100% |
| I/E | 4:1~1:8 |
| PEEP | 0cmH2O ~ 40 cmH2O |
| Pressure Limit | 20 cmH2O ~ 100 cmH2O |
| Monitored Parameters | |
| Frequency (Freq) | 0 /min ~ 100 /min |
| Tidal volume(Vt) | 0 mL ~ 2500 mL |
| MV
Airway pressure Dynamic lung compliance testing ETCO2 Concentration |
0 L/min ~ 99 L/min
0 cmH2O ~ 100 cmH2O 1mL/cmH2O ~ 1000mL/cmH2O 0mmHg ~ 152mmHg (0%-20%) |
| Oxygen concentration | 15 % ~ 100 % |
| Oscillographs display: | |
| P-T(pressure – time) | |
| F-T(flow – time) | |
| P-V loop (pressure – volume loop) | |
| V-T(volume – time ) | |
| ETCO2-T (End-tidal CO2 – time) | |
| Alarms and protection | |
| The AC power failure alarm | Power failure or no connection |
| Internal battery backup low voltage alarm | ≤11.3±0.3V |
| No tidal volume | No tidal volume within 6s |
| High Minute Volume alarm
Low Minute Volume alarm |
5L/min-99L/min
1L/min-30L/min |
| High Airway pressure alarm
Low Airway pressure alarm High oxygen concentration alarm Low oxygen concentration alarm |
20cmH2O-100cmH2O
0cmH2O-20cmH2O 19%-100% 18%-99% |
| Continuous pressure alarm | (PEEP+1.5cmH2O) over 16s |
| Suffocation warning
Fan error Oxygen deficit |
5-60s
Show on screen Show on screen |
| The maximum limited pressure | <12.5 kPa |
| Working conditions | |
| Gas source | O2, AIR |
| Pressure | 280kPa-600kPa |
| Voltage | ~220V±22V |
| Power frequency | 50Hz ± 1Hz |
| Input power | 900VA (with air compressor)
250VA (without air compressor) |
| Ventilation modes: | IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV |
| Ventilator parameter ranges | |
| Tidal volume(Vt) | 0 ~ 2000mL |
| Frequence (Freq)
Oxygen concentration |
1bpm ~ 100 bpm
21%-100% |
| I/E | 4:1~1:8 |
| PEEP | 0cmH2O ~ 40 cmH2O |
| Pressure Limit | 20 cmH2O ~ 100 cmH2O |
| Monitored Parameters | |
| Frequence (Freq) | 0 /min ~ 100 /min |
| Tidal volume(Vt) | 0 mL ~ 2500 mL |
| MV
Airway pressure Dynamic lung compliance testing ETCO2 Concentration |
0 L/min ~ 99 L/min
0 cmH2O ~ 100 cmH2O 1mL/cmH2O ~ 1000mL/cmH2O 0mmHg ~ 152mmHg (0%-20%) |
| Oxygen concentration | 15 % ~ 100 % |
| Graphical display: | |
| P-T(pressure – time) | |
| F-T(flow – time) | |
| P-V loop (pressure – volume loop) | |
| V-T(volume – time ) | |
| ETCO2-T (End-tidal CO2 – time) | |
| Alarm and protection | |
| The AC power failure alarm | Power failure or no connection |
| Internal battery backup low voltage alarm | ≤11.3±0.3V |
| No tidal volume | No tidal volume within 6s |
| High Minute Volume alarm
Low Minute Volume alarm |
5L/min-99L/min
1L/min-30L/min |
| High Airway pressure alarm
Low Airway pressure alarm High oxygen concentration alarm Low oxygen concentration alarm |
20cmH2O-100cmH2O
0cmH2O-20cmH2O 19%-100% 18%-99% |
| Continuous pressure alarm | (PEEP+1.5cmH2O) over 16s |
| Suffocation warning
Fan error Oxygen deficit |
5-60s
Show on screen Show on screen |
| The maximum limited pressure | <12.5 kPa |
| Working conditions | |
| Gas source | O2,AIR |
| Pressure | 280kPa-600kPa |
| Voltage | ~220V ±22V |
| Power frequency | 50Hz ± 1Hz |
| Input power | 900VA (with air compressor)
250VA (without air compreesor) |